48
Operator’s Manual Simple AFG Application
Berkeley Nucleonics Corporation
[email protected] | www.berkeleynucleonics.com
2955 Kerner Blvd, San Rafael, CA 94901 | 800-234-7858
Rev 1.1
Reference Clock Output Value
Connector
1 SMA on the Rear Panel
Output impedance
50
Ω
, AC coupled
Output Frequency
10MHz (100 MHz optional)
External Clock Input Connector
This connector input gives the User the possibility to feed directly a sampling clock for the System. This
Clock bypasses the internal Clock Synthesizer of the instrument.
The connector type is a SMA.
External Clock Input
Value
Connector type
1 SMA on the Rear Panel
Input impedance
50 Ohm, AC coupled
Input Frequency range
192.5 MHz, 385 MHz, 770 MHz or 1540 MHz (selectable)
Sync Clock Output Connector
This connector outputs a divided clock generated from sampling clock of the instrument. The resulting
frequency can be selected by the User from a list of all possible values.
The connector type is a SMA.
Sync Clock Output
Value
Connector type
1 SMA on the Rear Panel
Input impedance
50 Ohm, AC coupled
Input Frequency range
6.16Ghz / N where N=16, 32, 64, …, 2048
Digital Output connector
Important Note
: this connector is not used by AFG application.
Sync In / Sync Out Connectors
The purpose of those connectors is to connect and synchronize together multiple instruments: up to 4
instruments can be linked together.
Important Note
: this connector is not used by AFG application.