background image

InH2

®

 

Heated Humidifier

 User Manual V2.7

 

11/13 

English 

Guidance and manufacturer's declaration - electromagnetic immunity

 

The RESmart

®

 is intended for use in the electromagnetic environment specified 

below. The user of the RESmart

®

 should make sure that it is used in such an 

environment.  

Immunity 
test

 

IEC 60601 
test level

 

Compliance 
level

 

Electromagnetic 
environment - guidance 

 

 
Conducted 
RF 
IEC 
61000-4-6 
 
 
Radiated RF 
IEC 
61000-4-3 

 
3 Vrms 
150 kHz to 80 
MHz 
 
 
 
3 V/m 
80 MHz to 2.5 
GHz 

3 Vrms 

 
 
 
 
 
3 V/m 

Portable and mobile RF 
communications equipment 
should be used no closer to any 
part of the RESmart

®

, including 

cables, than the recommended 
separation distance calculated 
from the equation applicable to 
the frequency of the transmitter.

 

Recommended separation 
distance 
 

d

=1.2

d

=1.2

P    

80 MHz to 800 

MHz 

d

=2.3

P    

800 MHz to 2.5 

GHz 
 
Where 

P

 is the maximum output 

power rating of the transmitter 
in watts (W) according to the 
transmitter manufacturer and 

d

 

is the recommended separation 
distance in meters (m). 

 
Field strengths from fixed RF 
transmitter, as determined by an 
electromagnetic site survey, 

a

 

should be less than the 
compliance level in each 
frequency range. 

Interference 

may occur in the vicinity of 
equipment marked with the 
following symbol:

 

Summary of Contents for INH2

Page 1: ...English InH2 Heated Humidifier User Manual 0123...

Page 2: ...er is only used for single patient and must not be re used on another person This is to avoid the risk of cross infection The InH2 Heated Humidifier is not intended for use with a patient whose upper...

Page 3: ...only accessories recommended by BMC Never operate the humidifier if any of the parts are damaged if it is not working properly or if the humidifier has been dropped or mishandled Do not use the humidi...

Page 4: ...indicates the maximum water level for safe operation Outlet Port Connect the flexible tubing coaxial 22 mm here Water Chamber The removable water chamber holds the water for humidification It has a s...

Page 5: ...ector Plug this connector into the power outlet on the BMC RESmart Device Set Up 1 Connect the Humidifier to the RESmart device Make sure they are connected completely 2 Place the RESmart device and h...

Page 6: ...ifier will work automatically The yellow indicator light on the humidifier will turn on Press the humidifier button can turn off or restart the humidifier 5 The ideal humidity setting depends on room...

Page 7: ...mes with clean water and allow to air dry 4 Fill the water chamber and close the silica gel cover Inspect the water chamber for any leak or damage Replace the water chamber if any damage is present Hu...

Page 8: ...condensing Atmospheric Pressure 860 to 1060 hPa 500 to 1060 hPa Power Requirements 24V DC 1 0A max Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Sh...

Page 9: ...the System When traveling the optional CPAP carrying case is for carry on luggage only The carrying case will not protect the humidifier if it is put through checked baggage Security Stations For eas...

Page 10: ...missions CISPR 11 Group 1 Class B The RESmart uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Page 11: ...for power supply lines 1 kV for input output lines Mains power quality should be that of a typical home or hospital Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2...

Page 12: ...re than is indicated in the device specification it may be necessary to position the RESmart further from sources of power frequency magnetic fields The power frequency magnetic field should be measur...

Page 13: ...nications equipment should be used no closer to any part of the RESmart including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transm...

Page 14: ...adio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey sh...

Page 15: ...ive material or part BMC Medical Co Ltd will pay customary freight charges from BMC Medical Co Ltd to the dealer location only This warranty does not cover damage caused by accident misuse abuse alter...

Reviews: