BPAP System User Manual
41 / 57
17.6 Reposition of Alarming
After the alarm faults are cleared, the residual alarm messages still remain (alarm messages
are shown on the top of the screen without any visual and auditory alarm). Turn
the Knob
leftwards or rightwards to reduce the residual alarm messages.
17.7 Alarm Journal
The alarm journal is designed to record the latest 6 alarm messages. Retained inside the
device, the alarm journal will not be lost after a power interruption and the latest alarm
messages will replace the previous one, retaining 6 messages.
WARNINGS!
• Before using the device, the oeprator should check whether the current alarm preset value
is applicable to each patient. Such preset value can only be changed by a medical
professional and can not be modified by the patients at home
that a potential hazard can
exist if different ALARM PRE-SETS are used.
• In the case of a power failure or power cut for no more than 30 seconds, the last set alarm
value will be restored at the next operation.
CAUTION!
• The message in the alarming journal will be maintained when the device is powered down,
but the instantaneous time of power down will not be recorded.
17.8 Alarming Verification
Turn on the device, and then check the alarm system of the device at any time.
Tube disconnect alarm test
(1) When the device is operating normally, adjust the device to the appropriate patient
settings. Disconnect the tube that is connected to the air outlet of the device, and then verify
whether the tube disconnect alarm occurs.
(2) Press the home button
and it will become silent for 100 to 120 seconds. If the
alarm state is not eliminated, the alarm will sound again immediately at the end of the
silence.
(3) Reinstall the tube.
(4) Turn the button
leftwards or rightwards to reduce the residual alarm message.
Mask Blocked alarm test
(1) When the device is operating normally, adjust the device to the appropriate patient
settings. Block the vent hole of the mask for 35 seconds by hand or soft cloth, and then verify
whether the mask blocked alarm occurs.
(2) Press the home button
and it will become silent for 100 to 120 seconds. If the
alarm state is not eliminated, the alarm will sound again immediately at the end of the
silence.
(3) Turn the button
leftwards or rightwards to reduce the residual alarm message.