BPAP System User Manual
35 / 57
16.2 Options in the Patient Menu and Corresponding
Descriptions
Option
Range
Description
Humidifier
Off, Auto,
1
~
5
There are six humidity levels available. As the
number increases, the humidity rises accordingly.
―Off‖ means the humidifier is turned off.
Preheat
On / Off
Set humidifier to preheat by adjusting this option.
This feature is automatically turned off after 30
minutes
Reslex
Off, 1
~
3
This feature enables the device to automatically
reduce the treatment pressure when the patient
exhales, so as to make the patient more
comfortable. The higher the number, the more
pressure the device reduces. ―Off‖ means this
feature is disabled.
Heated Tubing
Off, 1
~
5
There are five heat levels available. As the number
increases, the heat rises accordingly. ―Off‖ means
the heat is turned off.
Note:
Heated Tubing
is displayed in the patient
menu only when a heated tubing is connected.
Ramp Time
Auto,
0
~
Max
Ramp
In order to increase comfort and help the patient fall
asleep easily, the pressure can be increased
gradually, when the Ramp feature is enabled. The
ramp time during which the initial pressure rises to
the preset treatment pressure can be adjusted. As
you turn
the Knob
to the nearest point, the
number increases or decreases by five seconds. The
screen displays a real-time countdown of the
remaining ramp time in seconds.
Delay Off
On / Off
When the humidifier is on, this feature allows the
airflow to continue for about 15 minutes at a low
pressure (about 2 hPa) after you press
the Start /
Stop button
to discontinue the treatment.
In this process, the vapor left in the water chamber
will be blown away to avoid any damage to the
device. When this feature is set to ―Off‖, which
means it is disabled, the device will stop delivering
air instantly after you press
the Start / Stop
button
.
Date
2000-01-01
—
2099-12-31
Set date by adjusting this option.
Date Format
yy mm dd /
mm dd yy /
dd mm yy
Turn
the Knob
to choose among three date
formats.
Time
00:00
—
23:59
Set time by adjusting this option.
Summary of Contents for G3 B20A