BPAP System User Manual
25 / 57
Fig. 13-6
WARNING!
• For safety, the device must be placed on a flat surface below the height of the patient's
head when he is lying on a bed, so that the condensation flows back to the water chamber
rather than remaining in the tubing which can cause droplet spraying.
CAUTIONS!
• Avoid moving or tilting the device when the water chamber has water in it.
• Take precautions to protect furniture from water damage.
13.2 Emptying the Water Chamber
(1)
Removing the water chamber
according to instructions in 13.1.1.
(2)
Emptying the water chamber:
Open the cap, as
shown in Fig. 13-7
, and pour any
remaining water out of the water chamber.
Fig. 13-7
CAUTION!
• Empty and air-dry the water chamber when the device is not in use.
(3)
Putting the water chamber
back
according to instructions in 13.1.3.
13.3 Setting the Humidity Level
After the device is powered on, turn
the Knob
to turn the heated humidifier on or off
and to adjust the humidity level according to instructions in the Patient Menu of the device.
There are five humidity levels available, and the number of humidity level will appear on the
screen of the device. The number 2 next to the icon indicates that the humidity is
adjusted to Level 2, as shown in Fig. 13-8. The water temperature in the water chamber is
maintained at a constant set level.