7
7
Fitting the Valve
Do not overtighten the locking pin.
We recommend using a locking pin that has a shoulder of between 13.5 mm and 19 mm in
diameter.
Valve
Locking Pin
Shoulder
Washer
Spacer
Pin Lock
Fit the spacer between the pin lock and the
valve only if the pin lock prevents the valve
from opening.
8
Fitting the Spacer
1
.
2
.
3
.
4
.
938399PK1/1-0521
Put the washer between the locking pin and the
valve. Apply Loctite 243 to the threads of the
locking pin.
Put the valve on the distal cap.
Confirm sufficient clearance in socket for valve to
operate correctly.
Torque the locking pin to 3 Nm, or tighten it with
your fingers by either 1/8 to 1/4 of a turn.
Summary of Contents for Silcare Breathe SBTTLP Series
Page 83: ...83 938399PK1 1 0521...