Voltage Calibration
Follow the step by step instructions below for voltage calibration.
1.
Power off the supply and connect its output terminal to a Digital
Volt Meter (DVM), as shown in the figure below. Then set the
meter for DC voltage measurement.
DVM
DC Voltage measurment
Voltage Calibration
DS30052
+
-
2.
Turn on the power supply’s output.
3.
Press
in the third page of Menu Setting to enter
CALIBRATION page. Users can enter the password and then access to
the calibration procedures.