background image

Subject to change without notice

10

set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15625x10

-5

) = 6.4div.

Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x10

3

) = 0.25ms/div,

min. time coefficient Tc = 1:(10x10

3

) = 0.1ms/div,

set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(10

3

x0.2x10

-3

) = 5div.

Displayed wavelength L = 0.8div,
set time coefficient Tc = 0.5µs/div,
pressed X-MAG. (x10) pushbutton: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10

-6

) = 25MHz,

required period T = 1:(25x10

6

) = 40ns.

If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be
applied (X-MAG. (x10) active). In this case, the time interval
of interest can be shifted to the screen center using the X-
POS. control.

When investigating pulse or square waveforms, the critical
feature is the rise time of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly
influence the measuring accuracy, the rise time is generally
measured between 10% and 90% of the vertical pulse height.
For measurement, adjust the Y deflection coefficient using
its variable function (uncalibrated) together with the Y-POS.
control so that the pulse height is precisely aligned with the
0% and 100% lines of the internal graticule. The 10% and
90% points of the signal will now coincide with the 10% and
90% graticule lines. The rise time is given by the product of
the horizontal distance in div between these two coincident
points and the calibrated time coefficient setting. The fall
time of a pulse can also be measured by using this method.

The following figure shows correct positioning of the
oscilloscope trace for accurate rise time measurement.

With a time coefficient of 5ns/div (X x10 magnification active),
the example shown in the above figure results in a total
measured rise time of

t

tot

 = 1.6div x 5ns/div : 10 = 

8ns

When very fast rise times are being measured, the rise times
of the oscilloscope amplifier and of the attenuator probe has
to be deducted from the measured time value. The rise time
of the signal can be calculated using the following formula.

In this t

tot

 is the total measured rise time, t

osc

 is the rise time

of the oscilloscope amplifier (approx. 2.3ns), and t

p

 the rise

time of the probe (e.g. = 2ns). If t

tot

 is greater than 34ns,

then t

tot

 can be taken as the rise time of the pulse, and calcu-

lation is unnecessary.

Calculation of the example in the figure above results in a
signal rise time

The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to
measure in any display position and at any signal amplitude.
It is only important that the full height of the signal edge of
interest is visible in its full length at not too great steepness
and that the horizontal distance at 10% and 90% of the
amplitude is measured. If the edge shows rounding or over-
shooting, the 100% should not be related to the peak values
but to the mean pulse heights. Breaks or peaks (glitches)
next to the edge are also not taken into account. With very
severe transient distortions, the rise and fall time measure-
ment has little meaning. For amplifiers with approximately
constant group delay (therefore good pulse transmission
performance) the following numerical relationship between
rise time

 tr (in ns)

 and bandwidth 

B (in MHz)

 applies:

Connection of Test Signal

In most cases briefly depressing the

 AUTO SET 

causes a

useful signal related instrument setting. The following
explanations refer to special applications and/or signals,
demanding a manual instrument setting. The description of
the controls is explained in the section

 “controls and read-

out”

.

Caution:
When connecting unknown signals to the oscil-
loscope input, always use automatic triggering and
set the input coupling switch to AC (readout). The
attenuator should initially be set to 20V/div.

Sometimes the trace will disappear after an input signal has
been applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is
only 3-8div. With a signal amplitude greater than 160Vpp and
the deflection coefficient 

(VOLTS/DIV.)

 in calibrated

condition, an attenuator probe must be inserted before the
vertical input. If, after applying the signal, the trace is nearly
blanked, the period of the signal is probably substantially
longer than the set time deflection coefficient (

TIME/DIV.

).

It should be switched to an adequately larger time coefficient.

The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as
PR37AG,  or  reduced  through  a  x10 o r x100  attenuator
probe. The use of test cables with high impedance circuits is
only recommended for relatively low frequencies (up to
approx. 50kHz). For higher frequencies, the signal source
must be of low impedance, i.e. matched to the characteristic
resistance of the cable (as a rule 50

). Especially when

transmitting square and pulse signals, a resistor equal to the
characteristic impedance of the cable must also be connected
across the cable directly at the Y-input of the oscilloscope.
When using a 50 

 cable.  When

transmitting square signals with short rise times, transient
phenomena on the edges and top of the signal may become
visible if the correct termination is not used. A terminating
resistance is sometimes recommended with sine signals as
well. Certain amplifiers, generators or their attenuators
maintain the nominal output voltage independent of frequency
only if their connection cable is terminated with the prescribed
resistance. If a

Type of signal voltage

Summary of Contents for 5105A

Page 1: ...150 MHz 200MS s ANALOG DIGITAL OSCILLOSCOPE INSTRUCTION MANUAL MODELS 5105A ...

Page 2: ...nal triggering 31 HOLD OFF time adjustment 31 B time base 2nd time base 31 Triggering after Delay 31 AUTO SET 32 Component Tester analog mode 32 General 32 Using the Component Tester 33 Test Procedure 33 Test Pattern Displays 33 Testing Resistors 33 Testing Capacitors and Inductors 33 Testing Semiconductors 33 Testing Diodes 33 Testing Transistors 34 In Circuit Tests 34 Storage mode 34 Signal reco...

Page 3: ... meters long If an interface has several connectors only one connector must have a connection to a cable Basically interconnections must have a double shielding 2 Signal cables Basically test leads for signal interconnection between test point and instrument should be as short as possible Without instruction in the manual for a shorter length signal lines must be less than 3 meters long Signal lin...

Page 4: ......

Page 5: ...only be inserted in a socket outlet provided with a protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor The main line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and the main line supply contacts line live neutral of the instrument have b...

Page 6: ...r supply Size 5x20mm 0 8A 250V AC fuse must meet IEC specification 127 Sheet III or DIN 41 662 or DIN 41 571 sheet 3 Time characteristic fast F This fuse must not be replaced by the operator should be kept in a clean and dry room and must not be operated in explosive corrosive dusty or moist environ ments The oscilloscope can be operated in any position but the convection cooling must not be impai...

Page 7: ...wards at each change Pure direct voltages can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean square value However for signa...

Page 8: ...n this case an appropriate high voltage capacitor approx 22 68nF must be connected in series with the input tip of the probe With Y POS control input coupling to GD it is possible to use a horizontal graticule line as reference line for ground potential before the measurement It can lie below or above the horizontal central line according to whether positive and or negative deviations from the gro...

Page 9: ...at the full height of the signal edge of interest is visible in its full length at not too great steepness and that the horizontal distance at 10 and 90 of the amplitude is measured If the edge shows rounding or over shooting the 100 should not be related to the peak values but to the mean pulse heights Breaks or peaks glitches next to the edge are also not taken into account With very severe tran...

Page 10: ...ing problems are eliminated Hum or interference appearing in the measuring circuit especially when a small deflection coefficient is used is possibly caused by multiple grounding because equalizing currents can flow in the shielding of the test cables voltage drop between the protective conductor connections caused by external equipment connected to the mains line e g signal generators with interf...

Page 11: ... and no other LED on the front panel is lit except A All INTENS settings are stored after the instrument is switched off The AUTO SET function switches the readout on and selects A time base mode A LED lit The INTENS setting for each function is automatically set to the mean value if less intensity was previously selected 5 TR The trace rotation control can be adjusted with a small screwdriver ple...

Page 12: ...ally again it will be required to stop rotating the knob for at least about 2 seconds Attention The dynamic range limits of the A D converter may become visible if a Y position shift is performed after storage This can affect those signal parts which were originally above or below the screen 8 PTR PK Det Pushbutton with two functions Neither function is available in analog mode PTR Briefly pressin...

Page 13: ... STOR MODE Pushbuttons with associated LEDs These functions are not available in analog mode If digital SINGLE SGL mode has not been chosen one of the associated LEDs is lit The signal capture and display mode can be selected by pressing one of the pushbuttons The mode setting is indicated by one of the LEDs RFR ENV AVM and ROL and also displayed by the readout The only exception is in XY storage ...

Page 14: ...ng mode on or off SINGLE mode is indicated by the associated SGL LED SINGLE mode is available in digital as well as in analog mode and remains unchanged when switching over from analog to digital mode or vice versa The main purpose of SINGLE is the capture of one time events but it can also be used in combination with repetitive signals SINGLE mode automatically selects A time base mode and normal...

Page 15: ...utton until a beep confirms reference memory acquisition Before starting the procedure the current signal can but must not be protected by the HOLD function As the reference signal is then equal to the current signal it is usually not visible at once If neither HOLD nor SINGLE is active the current signal can be shifted by the Y POS control s to make the reference signal visible If both REFERENCE ...

Page 16: ...o determine the DC content of a signal by comparing the actual Y position with the previously determined 0 Volt Y position Y POS I Symbol Provided that the readout is displayed and DC REFE RENCE ON is selected in the SETUP submenu MISCELLANEOUS the 0 Volt reference position is indicated by a ground symbol For channel I this symbol is displayed on the left of the vertical graticule center line if t...

Page 17: ...rning the control knob causes a different trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal voltage crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the symbol on the left vertical graticule line If the trigger point symbol would overwrite other readout information or would be invisible when being set ab...

Page 18: ... GD 34 38 Whether alternated or chopped channel switching is present in analog mode depends on the actual time base setting and is displayed in the readout ALT analog mode only displayed in the readout indicates alternate channel switching After each time base sweep the instrument internally switches over from channel I to channel II and vice versa This channel switching mode is automatically sele...

Page 19: ...te A B time base mode B time base mode and line mains trigger coupling STORAGE MODE ONLY In ROLL mode the signal capture is untriggered Con sequently alternate triggering is not available 25 VOLTS DIV Control knob This control knob for channel II has a double function The following description relates to the input attenuator function VAR LED dark Turning the control knob clockwise increases the se...

Page 20: ...start of the intensified sector In the free run condition delay trigger not active of the B time base an approximate value is displayed in the readout Dt This is an aid to find the position of the intensified sector which may be very small If only the B time base is being operated the delay time can be varied but there is no intensified sector as the A trace is not visible STORAGE MODE ONLY The HO...

Page 21: ... information is also displayed in the readout as an approximate value e g Dt 2 5ms related to the calibrated A time coefficient uncalibrated i e Dt 2 5ms The width of the segment decreases when the B time coefficient is set to a lower value higher time deflection speed Alternate time base mode causes the display to alter between A and B time base after each sweep analog mode In alternate DUAL mode...

Page 22: ...oefficient setting is not possible in storage mode Underneath the front panel sector described before the BNC sockets and four pushbuttons are located 32 INPUT CH I BNC socket This BNC socket is the signal input for channel I In XY mode signals at this input are used for the Y deflection The outer ground connection is galvanically connected to the instrument ground and consequently to the safety e...

Page 23: ... there are the controls for the readout the component tester and the square wave calibrator with their outputs Please note The following description of the cursor related controls assumes that the readout is visible and component test is inactive 40 PRINT MENU Pushbutton with 2 functions STORAGE MODE ONLY PRINT Briefly pressing the pushbutton starts a document hardcopy if the following preconditio...

Page 24: ...ust be activated Briefly pressing the CHI II 1 t ON OFF pushbutton switches from f to t and the following measure ment modes until the readout displays f again In all conditions vertical cursor lines are displayed even if voltage measurements are performed CX Time measurement of the active cursor is related to the trigger point If the active cursor position is on the left of the trigger symbol a n...

Page 25: ...ed in the readout must be multiplied by 100 if e g a x100 100 1 probe is used In case of x10 10 1 probes the probe factor can be automatically included see AC DC 33 and AC DC 37 t In time or frequency measurement condition two vertical cursor lines are displayed The measurement result is indicated as t time measurement or f frequency measurement Please note item 41 1 Dt NOTE For frequency measurem...

Page 26: ...evice under test The oscilloscope is switched on by depressing the red POWER pushbutton After a few seconds the B K logo and the instrument software release is displayed on the screen As long as the B K logo is visible different internal checks are made Thereafter the instrument will revert to its last used operating mode If after approx 20 seconds no trace is visible the AUTO SET pushbutton shoul...

Page 27: ...1MHz Operate the oscilloscope as described under 1kHz but select for 0 2µs div time deflection coefficient setting Insert the probe tip into the output socket A waveform will be displayed on the CRT screen with leading and trailing edges clearly visible For the HF adjustment now to be per formed it will be necessary to observe the rising edge as well as the upper left corner of the pulse top The l...

Page 28: ... are the same for both channels because the X x10 magnifier is inactive in XY mode The bandwidth of the X amplifier is lower than the Y amplifier and the phase angle which increases with higher frequencies must be taken into account please note data sheet The inversion of the X input signal is not possible Lissajous figures can be displayed in the X Y mode for certain measuring tasks Comparing two...

Page 29: ...that several cycles of the modulation frequency are visible Strictly speaking triggering should be external with modulation frequency from the AF generator or a demo dulator However internal triggering is frequently possible with normal triggering using a suitable trigger level setting and possibly also using the time vernier variable adjustment Figure 1 Amplitude and frequency spectrum for AM dis...

Page 30: ...gger coupling setting In the absence of an adequate trigger signal or when the trigger controls particularly the trigger LEVEL control are misadjusted no trace is visible i e the screen completely blanked When using the internal normal triggering mode it is possible to trigger at any amplitude point of a signal edge even with very complex signal shapes by adjusting the trigger LEVEL control If the...

Page 31: ...xternal trigger voltage itself is displayed first with internal triggering In most cases the composite video signal has a high DC content With constant video information e g test pattern or color bar generator the DC content can be suppressed easily by AC input coupling of the oscilloscope amplifier With a changing picture content e g normal program DC input coupling is recommended because the dis...

Page 32: ...aiting the trigger pulse This causes an untriggered display and a flashing trigger LED HOLD OFF time adjustment For instrument specific information please note DEL POS HO 28 in section Controls and Readout If it is found that a trigger point cannot be found on extremely complex signals even after careful adjustment of the trigger level control a stable display may often be obtained using the hold ...

Page 33: ...ly for a display of approx 2 signal periods The time base setting occurs randomly if complex signals consisting several frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was selected the last used input coupling AC or DC will be chosen Internal triggering cha...

Page 34: ...determination of actual values will come with experience or by direct comparison with a component of a known value Testing Capacitors and Inductors Capacitors and inductors cause a phase difference between current and voltage and therefore between the X and Y deflection giving an ellipse shaped display The position and opening width of the ellipse will vary according to the impedance value at 50Hz...

Page 35: ...ompared with single components In case of doubt one component terminal may be unsoldered This terminal should then not be connected to the ground socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which is known to be operational likewise without power and any external connections Using the test prods identical test points in each ci...

Page 36: ...nd Controls Vertical resolution The dot density in each operation mode is 8 bits 28 256 dots displayed over a height of roughly 10 divisions The instrument is adjusted for 25 dots per division This eases processing and cursor measurement Insignificant differences between the analog screen display and the digital data are unavoidable This concerns signal height as well as the position The trace pos...

Page 37: ...ack must on no never be visible Visible trace faults without input signal bright dot on the left side or decreasing brightness from left to right or shortening of the baseline Cause Incorrect Unblanking Pulse It should be noted that with wide variations in brightness refocusing is always necessary Moreover with maximum brightness no pumping of the dis play must occur If pumping does occur it is no...

Page 38: ...ting the trimmer located in the 2 1 attenuator The shape of the square wave should then be the same in each input attenuator position Operating Modes CH I II DUAL ADD CHOP INVERT and X Y Operation In DUAL mode two traces must appear immediately On actuation of the Y POS controls the trace positions should have minimal effect on each other Nevertheless this cannot be entirely avoided even in fully ...

Page 39: ... Then both readout information regarding the time deflection coefficient and the time difference show exactly the same value After switching over from time to frequency measurement the readout displays the required signal frequency in each time base setting Hold Off time Analog The variation of the hold off time setting can not be measured without opening the instrument However a visual check can ...

Page 40: ... the oscilloscope and the first command SPACE CR 20hex 0Dhex sent from the PC the baud rate is recognized and set automatically between 110 baud and 19200 baud The oscilloscope is then switched over to REMOTE control mode The oscilloscope RS232 Interface Remote Control then transmits the RETURNCODE 0 CR LF to the PC In this status all settings with the exception of those functions mentioned under ...

Page 41: ... 2001 B K Precision Corp 481 323 9 001 Printed in U S A 1031 Segovia Circle Placentia CA 92870 ...

Reviews: