Power Analysis
194
Figure 19.20
Connection Guide
19.8.2
Testing Conditions
Turn On
Determines the time taken for the power to reach a certain percentage of its steady-state output. The turning on time
is between T2 and T1, where:
•
T1
= When the input voltage first rises to a certain percentage (usually 10%) of its maximum amplitude
•
T2
= When the output DC voltage rises to a certain percentage (usually 90%) of its maximum amplitude
Turn Off
Defines the time that has elapsed after a power supply has been turned off and has fallen to a certain percentage of its
maximum output voltage. The turning off time is between T2 and T1, where:
•
T1
= When the input voltage finally falls to a certain percentage (usually 10%) of its maximum amplitude
•
T2
= When the output DC voltage finally falls to a certain percentage (usually 10%) of its maximum amplitude