11.5
Pass/Fail
One way to verify a waveform is within desired bounds is to use pass/fail testing. A pass/fail test
defines a region of the oscilloscope display in which the waveform must remain in order to pass
the test. Being within a specified-range is verified point-by-point across the display.
The test results can be displayed on the screen as well as be declared through the system sound
or a pulse output from the TRIG OUT connector on the rear panel. The Pass/Fail test only operates
on analog channels that are displayed.
Perform a Pass/Fail Test
1.
Press the Utility button to enter the UTILITY menu.
2.
Press the Next Page softkey to go to the second page of the UTILITY menu.
3.
Press the Pass/Fail softkey to enter the PASS/FAIL menu.
4.
Press the Enable Test softkey to show On to enable the pass/ fail test.
5.
Press the Source softkey to select the desired channel.
6.
Press the Mask Setting softkey to enter the MASK menu.
7.
Press the Mask softkey to select which parameter to adjust (X for the time axis, Y for the
voltage/current axis). Use the Universal Knob to adjust it to the desired value (larger values
mean wider acceptance intervals). The range is from 0.02 div to 4 div.
Figure 96 - Pass/Fail test