System Settings
120
14.6.1 Set and Perform a Pass/Fail Test
To set and perform pass/fail test
:
1. Press the
Utility
key on the front panel to enter the UTILITY function menu.
2. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
3. Press the
Pass/Fail
softkey to enter the PASS/FAIL function menu.
4. Press the
Enable Test
softkey to select On to enable the pass/ fail test.
5. Press the
Source
softkey to select the desired channel.
6. Press the
Mask Setting
softkey to enter the MASK function menu.
7. Press the
X Mask
or
Y Mask
softkey; and then turn the
Universal Knob
to select the desired value. The range is
from 0.02div to 4div.
8. Press the
Create Mask
softkey to create the mask. Whenever the
Create Mask
softkey is pressed the old mask is
erased and a new mask is created.
9. Press the
UP
softkey to return to the PASS/FAIL function menu.
10. Press the
Msg Display
to select
On
or
Off
to turn on or off the message display. When On is selected, the test
result will be displayed in the upper-right message box of the screen.
•
F
: the frame number of the channel waveform that fails to match the mask.
•
P
: the frame number of the channel waveform that matches the mask.
•
T
: the frame number of the channel waveform that have been acquired. It is the total of
Pass
and
Fail
.
11. Press the
Next Page
softkey to go to the second page of the PASS/FAIL function menu.
12. Press the
Stop On Fail
softkey to select
On
or
Off
to turn on or off the function.
•
On
: when failed waveforms are detected, the oscilloscope will stop the test and enter the STOP state. At this
point, the results of the test remain the same on the screen (if display is turned on) and only one pulse is output
from the
[TRIG OUT]
connector (if enabled) at the rear panel.
•
Off
: the oscilloscope will continue with the test even though failed waveforms are detected. The test results on
the screen will update continuously and a pulse will be output from the
[Trigger Out]
connector at the rear panel
each time a failed waveform is detected.
13. Press the
Output
softkey to select turn on or off the sound.
•
: When the failed waveforms are detected, the beeper sounds.
•
: The beeper will not sound even if the failed waveforms are detected,
14. Press the Next Page softkey to return to the first page of the PASS/FAIL function menu.
15. Press the Operate softkey to perform the test.
•
: Current state is stop; press the softkey to start the pass/fail test.
•
: Current state is running; press the softkey to stop the pass/fail test.
Summary of Contents for 2194
Page 1: ......
Page 82: ...Serial Trigger and Decode 82 Figure 7 29 LIN Trigger ...