Operating Instructions
10
Error Codes
The Bird 4421 displays error codes when the RF power is either below the selected range (underrange) or above the
selected range (overrange).
Table
1
displays the error codes and
Table
2
lists the function limits.
Table 1
Error
Codes
Table 2
Function
Limits
Audible Warning
If the RF power level exceeds 120
% of the power sensor’s maximum power capability, the power meter will sound a
warning buzzer.
Symbol
Explanation
Value greater than overrange limit of function
Value less than underrange limit of function
Function
Limit
Error
FWD, RFL
Power > 199.9
% of full scale or 120
%
of
top range
Overrange
FWD dBm, RFL dBm
Power > 120
% of full scale
Overrange
Power < 3
% of low range
Underrange
SWR
FWD < 20
% of low range
Underrange
FWD – RFL = 0
Overrange
Return Loss
FWD < 20
% of low range
Underrange
RFL < 20
% of low range
Underrange
Return Loss > 40 dB
Underrange