Recall a Saved Trace
When you recall a saved trace, it is displayed along
with the trace currently being measured (two traces on
the screen). To remove a recalled trace from the dis-
play, press the Save/Recall soft key then press the Esc
key. With a current and recalled trace, the display
behaves as follows:
•
The frequency range will change to the settings of
the recalled trace.
•
Recalled traces adopt the display type of the cur-
rent trace (envelope, maximum, minimum).
•
The resolution of the recalled trace does not affect
the resolution of the current trace.
•
Markers always follow the current trace not the
recalled trace.
•
When comparing the current trace to a recalled
trace, each data point on the current trace is com-
pared to only the first data point on the recalled
trace (at the frequency of the marker).
•
The status box at the top of the screen displays the
resolution and display type of the current trace on
the top line and the label text of the recalled trace
on the bottom line.
Note:
Calibration is automatically turned off if
the recalled trace has a different frequency range from
the trace currently being measured.
The recalled a trace is drawn using the same display
type (envelope, maximum, minimum) as the trace cur-
rently being measured.
You must be in either Measure Match mode or Fault
Location mode to recall a saved trace.
1. Press the Save/Recall soft key
2. Follow the numbered steps in Figure 51, page 105
to select and recall a saved trace.
Summary of Contents for SITE ANALYZER SA-1700EX
Page 2: ...This page is not blank...
Page 8: ...Bird Site Analyzer vi...
Page 9: ...vii...
Page 30: ...Bird Site Analyzer 16...
Page 142: ...Bird Site Analyzer 128...
Page 148: ...Bird Site Analyzer 134...
Page 159: ...Maintenance 145 Measured Return Loss dB...
Page 167: ...Maintenance 153 the number of data points you wish measure 238 475 949...