Measurements
125
Out-of-Band and In-Band, Out-of-Channel Spurious
NOTE
These are not automated measurement methods and are not listed on the measure menu.
Out-of-Band & In-Band, Out-of-Channel Spurious measures the distortion and interference
inside or outside a system band.
1.
Click on Mark & Limit.
2.
Click on the Select Marker: 1 2 3 4 5 6 option to select marker 1.
NOTE
The bracketed number indicates the active marker.
3.
Click on Marker: On/Off to turn on marker.
4.
Use the arrow keys, the number keys or mouse scroll wheel to move the marker over
one of the spurs.
5.
Compare the value of the marker to the specified allowable level of Out-of-Band (In-
Band, Out-of-Channel) spurious emissions for the corresponding channel transmit
frequency.