Wideband Power Sensors
28
Match Measurement
Measurement Range:
Return Loss
Rho (
)
VSWR
0 to 23 dB
0.07 to 1.0
1.15 to 99.9
Forward Power, Min
0.1 W
Measurement Uncert.
See
Figure
16 on page
28
.
Figure
16
Match Measure Uncertainty
-6
-4
-2
0
2
4
-25
-20
-15
-10
-5
0
Uncertai
n
ty
(d
B)
Return Loss
Measurement Uncertainty
Above
3GHz+
Above
3GHz-
Below
3GHz+
Below
3GHz-
Peak Envelope Power
RF Power Range
4 – 60 W
1
1 Max. power depends on frequency and system VSWR.
See
Figure
17 on page
29
.
Measurement Uncert.:
burst width > 200 μs
1 μs < b.w. < 200 μs
burst width < 1 μs
burst width < 0.5 μs
± (7% of r 0.05 W)
2
2 Above 35 °C or below 15 °C add 3%
For D < 0.1 add 0.1 W
For period > 0.1s add (1.5% + 0.15 W)
± (10% of r 0.1 W)
2
± (15% of r 0.1 W)
2
± (20% of r 0.1 W)
2
Minimum Pulse Width:
400 ns
Summary of Contents for 5012D
Page 24: ...Wideband Power Sensors 16 Figure 9 5012D WPS...
Page 29: ...Specifications 21 Figure 12 5016D WPS...
Page 34: ...Wideband Power Sensors 26 Figure 15 5017D WPS...
Page 39: ...Specifications 31 Figure 18 5018D WPS...
Page 44: ...Wideband Power Sensors 36 Figure 21 5019D WPS...
Page 45: ...Specifications 37...