5018D Specifications
Sensor Characteristics
Frequency Range
150 MHz to 4.0 GHz
RF Power Range
100 mW to 25 W average,
60 W peak
Maximum Power
See
Figure
17 on page
29
.
Impedance, Nominal
50 ohms
Insertion Loss, Max:
150 – 1000 MHz
1000 – 4000 MHz
0.05 dB
0.01 dB
Input VSWR, Max:
150 – 2500 MHz
2500 – 4000 MHz
1.05
1.10
Directivity, Min:
150 – 3000 MHz
3000 – 4000 MHz
30 dB
28 dB
RF Connectors
N Female
Interface:
DPM
PC Serial Port
PC USB Port
Male DB-9, RS-232, 9600 Baud, no
parity, 8 data bits, 1 stop bit
Female DB-9, EIA-232, 9600 Baud,
no parity, 8 data bits, 1 stop bit
USB 1.1 interface
Power Supply:
DPM
USB Port
DC Connector
From host instrument via cable
less than one low-power USB load
7 – 18 Vdc, < 100 mA
Data Logging
In VPM software
Average Power
RF Power Range
25 mW – 25 W
Peak/Average Ratio, Max
12 dB
Measurement Uncert.
± (4% of r 0.008 mW)
1
1 Above 35 °C or below 15 °C add 3%