Wideband Power Sensors
18
Match Measurement
Measurement Range:
Return Loss
Rho (
)
VSWR
0 to 23 dB
007 to 0.999
1.15 to 99.9
Forward Power, Min
0.1 W
Measurement Uncert.
See
Figure
10 on page
18
.
Figure
10
Match Measure Uncertainty
-6
-4
-2
0
2
4
-25
-20
-15
-10
-5
0
Uncertai
n
ty
(d
B)
Return Loss
Measurement Uncertainty
Above
3GHz+
Above
3GHz-
Below
3GHz+
Below
3GHz-
Peak Envelope Power
RF Power Range
1.0 W – 60 W
1
1 Max. power depends on frequency and system VSWR.
See
Figure
11 on page
19
.
Measurement Uncert.:
burst width > 200 μs
1 μs < b.w. < 200 μs
burst width < 1 μs
burst width < 0.5 μs
± (7% of r 0.05 W)
2
2 Above 35 °C or below 15 °C add 3%
For D < 0.1 add 0.1 W
For period > 0.1s add (1.5% + 0.15 W)
± (10% of r 0.1 W)
2
± (15% of r 0.1 W)
2
± (20% of r 0.1 W)
2
Minimum Pulse Width:
400 ns