Safety during Use
Disposal
6
Cleaning and Disinfection
Clean the device with a damp cloth and a mild soap solution with a pH value between 4 and 10.
For disinfecting, use an alcohol-based, hydrogen peroxide-based, or quaternary compound-based
disinfectant. The disinfection of the device was tested using the following agents:
•
Alcohol-based disinfectant
–
Mixture of 70% isopropyl and 30% water
–
Mixture of 85% ethanol and 15% water
–
Aerodesin 2000 (manufactured by Lysoform Dr. Hans Rosemann GmbH)
•
Hydrogen peroxide-based disinfectant
–
2-percent solution for use with terralin
®
PAA (manufactured by Schülke & Mayr GmbH)
•
Quaternary compound-based disinfectant
–
2-percent solution for use with terralin
®
protect (manufactured by Schülke & Mayr GmbH)
Sterilization
The device is non-sterile and must not be sterilized. Follow the instructions on cleaning and
disinfecting.
Disposal
The symbol on the type plate, a crossed out garbage can, indicates that the device must be disposed of
in accordance with Directive 2012/19/EU on waste electrical and electronic equipment (WEEE 2).
Disposal of this device in an environmentally unfriendly manner will result in environmental pollution,
as this device contains materials that must be disposed of in accordance with environmental protection
regulations (e.g., WEEE, RoHS, REACH). Dispose of the used product in an environmentally sound
manner and according to the applicable country-specific directives. Return products that are no longer
in use to BIOTRONIK.