Appendix
Package Contents
11
Category
Design
Model
Europe
Telit LE910C1‑EU
North America
Telit LE910C1‑NA
Asia-Pacific
Telit LE910C1‑AP
Latin America
Telit LE910C1‑LA
LTE bands in use
Europe
B1 (1,900 MHz), B3 (1,800 MHz), B7 (2,600 MHz),
B8 (900 MHz), B20 (800 MHz), B28A (700 MHz)
North America
B2 (1,900 MHz), B4 (1,700/2,100 MHz),
B12 (700 MHz)
Asia-Pacific
B1 (1,900 MHz), B3 (1,800 MHz), B5 (850 MHz),
B28 (700 MHz)
Latin America
B1 (2,100 MHz), B3 (1,900 MHz), B3 (1,800 MHz),
B4 (1,700/2,100 MHz), B7 (2,600 MHz),
B28 (700 MHz)
Package Contents
Information about the package contents can be found on the shipping label.
Electromagnetic Interferences
Possible electromagnetic interference
This device is protected against electromagnetic interference and electrostatic discharges in the
special environment in the vicinity of high-frequency surgical instruments. At the same time, the
emitted interference is reduced to a minimum. The device thus meets all requirements of
IEC 60601-1-2.
The following tests were conducted according to IEC 60601‑1‑2: 2014:
Section
Test
Test level
7.1
EN 55011 (CISPR 11)
Radiated emission
•
Group 1
•
Class B
8.9
IEC 61000‑4‑2
Electrostatic discharge (ESD)
•
± 8 kV contact discharge
•
± 15 kV air discharge
8.9/8.10
IEC 61000‑4‑3
Electromagnetic fields
•
Modulation: 1 kHz
•
3 V/m, 80 MHz – 2.7 GHz
•
Limits for RF communication
equipment per Table 9 in
IEC 60601-1-2 (9–28 V/m)