| Chapter
5
:
Instrument Qualification Process
Linearity Test
1. Calculate the Mean of the four wells for each concentration in columns 1–5.
2. Perform linear regression using these values as inputs:
Sodium Fluorescein
x
y
1000
Mean of the 1000 pM wells
500
Mean of the 500 pM wells
250
Mean of the 250 pM wells
125
Mean of the 125 pM wells
62.5
Mean of the 62.5 pM wells
Methylumbelliferone
x
y
100
Mean of the 100 nM wells
50
Mean of the 50 nM wells
25
Mean of the 25 nM wells
12.5
Mean of the 12.5 nM wells
6.25
Mean of the 6.25 nM wells
3. Calculate the R
2
value; it must be
>= 0.9500
to pass.
Troubleshooting
If any tests fail, please try the following suggestions. If the test(s) continue to fail, print the
results and contact
Technical Support
.
l
Are the solutions fresh? Discard the plate and any opened, unused test solutions after
seven days.
l
Are the excitation/emission filters clean?
l
Are you using the proper filter cube?