Chapter 5: As-Needed Maintenance
Clean the Tip Priming Trough and Priming Plate
1.
Remove the tip priming trough from the left rear of the instrument’s microplate
carrier (see below).
2.
Wash the tip priming trough and priming plate in hot, soapy water. Use a small
brush or cloth to clean the corners of the trough and plate.
3.
To decontaminate
,
soak the trough and plate in a container of 0.5% sodium
hypochlorite
or
70% isopropyl alcohol for 20 to 30 minutes.
4.
If decontaminating in bleach solution, remove the trough and plate, and
thoroughly rinse with DI water.
If decontaminating with alcohol, remove the trough and plate and let them air-
dry.
5.
Discard the used gloves and cloths using a biohazard trash bag and an
approved biohazard container.
Figure 33: Tip priming trough and priming plate
tip priming
trough
BioTek Instruments, Inc.