Flush/Purge Fluid Path
Applies only to models equipped with injectors
At the end of each day that the dispense module is in use, flush the fluid path using the
Gen5 priming utility. Leave the fluid to soak overnight or over a weekend, and then purge
the fluid before using the instrument again.
This flushing and purging routine is also recommended before disconnecting the outlet
tubes from the reader, and before decontamination to remove any assay residue prior
to applying isopropyl alcohol or sodium hypochlorite.
i
If using Gen5 version 2.05 or earlier, the
Prime
and
Purge
options are found under
the
Dispenser
tab in the Reader Control dialog.
To flush the fluid path:
1. Fill two supply bottles with deionized or distilled water. Insert the supply (inlet) tubes
into the bottles.
2. Place the priming plate on the carrier.
3. Select
System > Instrument Control > Synergy H1
.
4. Click the
Prime
tab and select Dispenser 1.
5. Set the Volume to 5000 µL. Keep the default prime rate.
6. Click
Prime
to start the process. When the process is complete, carefully remove
the priming plate from the carrier and empty it.
7. Repeat the process for Dispenser 2.
Leave the water in the system overnight or until the instrument will be used again.
Purge the fluid from the system (see below) and then prime with the dispense
reagent before running an assay.
To purge the fluid from the system:
1. Place the inlet tubes in empty supply bottles or a beaker.
2. Select
System > Instrument Control > Synergy H1
.
3. Click the
Prime
tab and select Dispenser 1.
4. Set the Volume to 2000 µL.
5. Click
Purge
to start the process.
6. When the purge is complete, repeat the process for Dispenser 2.
After purging the system, you may wish to run a quick Dispense protocol to visually
verify the dispense accuracy (see the next section) or the more thorough Dispense
Accuracy and Precision Tests (see
starting on page
).
Synergy H1 Operator's Manual
Flush/Purge Fluid Path | 61
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...