Select Band Pass, Long Pass, or Short Pass and enter the wavelength and
bandwidth.
l
Select Plug to indicate the presence of a plug.
l
Select Hole to indicate an empty location.
6. Select the mirror type and enter the excitation and emission ranges. Note that the
“M” value on the filter cube label is the cut-off (nm).
Cut-off (nm)
Excitation Range
Emission Range
50%
200-850
200-850
320
260-305
335-750
365
290-350
380-800
400
320-390
410-800
435
385-425
445-610
455
400-450
460-710
510
440-505
515-640
525
475-520
530-670
545
512-535
555-578
550
415-540
560-850
555
541-550
560-595
570
515-565
575-735
595
540-590
600-770
635
640-780
400-630
660
580-655
665-850
i
If 'Fluorescence Polarization Cube' is checked, only Filter Set 1 is
available for definition. The filters and mirrors of Filter Set 2 must be
identical to those of Filter Set 1 for fluorescence polarization.
7. Define Filter Set 2, if necessary.
8. Click
OK
to return to the Filter Cubes dialog. Click the Help button to learn about the
Read Plate Prompt options.
48 | Chapter 4: Filters and Mirrors