13: Run a System Test
Running a System Test will confirm that the reader is set up and operating properly, or will
provide an error code if a problem is detected.
1. Turn on the incubator:
l
In Gen5, select
System > Instrument Control > Synergy H1
.
l
Click the
Pre-Heating
tab. Enter a Requested temperature of at least 37°C and
then click
On
.
Wait until the incubator temperature reaches the set point before continuing.
2. Select
System > Diagnostics > Run System Test
. Select your reader if
prompted and click
OK
.
3. When the test is complete, a dialog requesting additional information appears. Enter
the information and click
OK
.
If a message appears stating that the reader has a
pending
system test report,
view the report and then repeat steps 2 and 3.
4. The results report appears and should contain the text “SYSTEM TEST PASS”.
l
If required, print the report and store it with your installation records. Note: Gen5
stores results in its database; you can print a report at any time.
If an error code is returned, refer to
starting on page
. If the
problem is something you can fix, do so now and run another System Test. If
the problem is something you cannot fix, or if the test continues to fail, contact
BioTek’s Technical Assistance Center.
5. Turn off the incubator.
Models with injectors:
Keep Gen5 open and proceed to the next section.
All other models:
The installation and setup process is complete. Close Gen5 and proceed
to
Operational/Performance Qualification
on page
Synergy H1 Operator's Manual
13: Run a System Test | 19
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...