7: Install the Dispense Module
Applies only to models equipped with injectors
1. Place the dispense module on top of the reader or, if equipped, on top of the gas
controller module.
i
Do not place the dispense module
next to
the reader.
Figure 2-2: Dispense module on top of the reader (shown without
the gas controller)
2. Open the bag containing the injector tubes and tips. Remove the clear shrouds from
the tubes.
3. Remove the two inlet tubes from their canisters.
4. Identify the two syringe valves on the dispense module (see
on page
).
Each is labeled with a left-pointing arrow.
i
When installing the tubes, do not use any tools. Finger-tighten only!
5. Screw the fitting of one inlet tube into the right side of the Syringe 1 valve.
6. Screw one end of one outlet tube into the left side of the Syringe 1 valve.
7. Repeat these steps to attach the inlet and outlet tubing for Syringe 2.
Synergy H1 Operator's Manual
7: Install the Dispense Module | 13