Luminescence Testing
For models with luminescence capability, BioTek uses the Harta Luminometer Reference
Microplate to test the luminescence system. The test plate is LED-based and NIST-
traceable. Contact BioTek to purchase a plate (BTI #8030015; includes microplate carrier
adapters) or visit www.HartaInstruments.com to learn more.
Test Method
The Harta Luminometer Reference Microplate is used to determine a detection limit by
leveraging a known correlation of 35 photons per attomole of ATP. By using the NIST data
provided with the Harta plate in photons/s, a conversion factor of 0.02884 attomole/photon
is applied to determine an ATP concentration and subsequent limit of detection for the
instrument under test.
Gen5 Protocol Parameters
The information in this section represents the recommended reading parameters for the
referenced Gen5 protocol(s).
Synergy H1 F-LumTest_Harta.prt
Parameter
Setting
Plate Type
If present "8030015 Harta - w/o 8032028
adapter" otherwise "Costar 96 black opaque"
Delay Step
3 minutes
READ STEP 1
Detection Method
Luminescence
Read Type
Endpoint
Optics Type
Filters
Step Label
Reference well A2
Read well
A2
Filter Set
1 (filter cube)
Excitation
Plug
Emission
Hole
Gain
135
Integration Time
0:10.00 MM:SS.ss
Delay After Plate Movement
350 msec
Dynamic Range
Standard
Read Height
7.00 mm
Synergy H1 Operator's Manual
Luminescence Testing | 83
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...