Instrument System Test
Each time the Synergy H1 is turned on, it automatically performs a series of tests on the
reader’s motors, lamp(s), the PMT, and various sub-systems. The duration of this “system
test” depends on the reader model and can take a few minutes to complete. If all tests
pass, the microplate carrier will eject and the LED on the power switch will remain on and
constant. The reader is then ready for use.
If any test results do not meet the internally coded Failure Mode Effects Analysis (FMEA)
criteria established by BioTek, the reader will beep repeatedly and the LED on the power
switch will flash. If this occurs, press the carrier eject button to stop the beeping. If
necessary, initiate another system test using Gen5 to try to retrieve an error code from the
reader.
Refer to
starting on page
for information on error codes and
troubleshooting tips.
Refer to
on page
to see a sample System Test Report for Synergy
H1.
Plate Shaker Test
This test verifies that the multi-speed plate shaker is operating properly. The test involves
creating and running a protocol with shaking enabled for a duration of 30 seconds. The
sound of the carrier shaking is all that needs to be confirmed to verify that the plate shaker
is operating properly.
74 | Chapter 7: Instrument Qualification Process
BioTek Instruments, Inc.
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...