Note the flow direction arrows on the check valves. Some are harder to see than
others. The flow direction arrows are engraved valves into the translucent plastic.
l
PN 68083 — Autoclavable valves for use with non-organic substances. The direction
arrows are difficult to see.
l
PN 68073 — Check valves recommended for use with organic substances. They cannot be
autoclaved. Direction arrows are easy to see.
n
Note: If the check valves are replaced, it may be necessary to recalibrate the
syringe backlash to achieve optimum accuracy performance.
the Backlash for Syringe Dispenser on page 212
Make sure the flow-direction arrows on the check valves point toward the pump
from the supply vessel, and towards the dispense manifold from the pump.
Install Inline Filter for 32-Tube Dispensers
BioTek ships two 90-micron inline filters with the 32-tube dispense manifolds to
reduce the chances of clogging the dispense tubes. It is especially important for the
SB – small bore models.
To install the filters:
1. Locate and layout the length of tubing that goes between the supply bottle and
the Syringe pump; it has the Luer fitting on one end.
2. Cut the tubing approximately one to two inches (3-5 cm) above the Luer fitting.
3. Slide the filter’s nozzle ends into the two ends of the tubing, reconnecting it.
Install Software/Connect to Computer
If you purchased BioTek's Liquid Handling Control™ (LHC) Software to control the
MultiFlo FX using your personal computer (PC), please refer to the LHC
Installation Guide for complete installation and setup instructions.