294 | Appendix A: Error Codes
MultiFlo FX Software Error Codes
Generally, these errors are caused by protocol parameters that conflict with the
instrument's onboard settings. The protocol may have been originally created for a
different hardware configuration, the 192-tube wash manifold instead of the 96-
tube, for example.
Error
Code
Description
Help
6000
General communication error during
download.
See
6001
COM port created by USB converter
no longer active.
See
6002
Invalid basecode part number;
instrument is not MultiFlo FX.
Service Only.
6003
Invalid Basecode Data Version;
basecode needs to be updated.
Contact BioTek to obtain latest
basecode.
6004
No rows are selected for the specified
plate type.
Modify the protocol to select a row.
6005
Invalid row selection value (must be
0 or 1).
Contact BioTek TAC.
6006
This instrument can only process 96-
well plates.
The protocol may have been created for
another instrument, change the plate
type or select another protocol.
6007
This instrument can only process
1536-well plates.
6008
The 8-tube Syringe Manifold can only
be used with 96-well plates.
Mismatch between installed hardware
and protocol parameters: change the
plate type or correct the instrument
settings to match the currently installed
hardware.
6009
The 96-tube single wash manifold can
only be used with 96-well plates.
6010
The data is invalid or out-of-range.
Service Only.
Contact BioTek TAC.
6011
This step type cannot be
downloaded.
Review the limitations to transferring
protocols to the instrument, See the
LHC Help Topic:
Transferring Protocols
.
MultiFlo™ FX Multi-Mode Dispenser