200 | Chapter 4: Maintenance
With strict adherence to best practices and maintenance recommendations, this is
the typical longevity of the dispense cassettes.
Example table for recording cassette usage:
You may want to create a form similar to this table to keep track of the volume
dispensed with each cassette:
Cassette serial #:
2178
Date
# Plates Plate Type Volume/Well Total Daily Vol. Total Cassette Vol
10/10/08
26
384
5 µL
49920 µL
50 mL
10/11/08
33
96
10 µL
31680 µL
82 mL
RAD 8-to-1 Cassette - Replace Chute
As needed, you can replace the chute on the 8-tube RAD bulk-dispensing cassette.
A spare chute is included in the cassette package.
Use the Philips screwdriver shipped with the MultiFlo FX to remove the chute and
replace it with another:
To reattach the chute, align it on the pins on the tip head and secure it with its two
screws. Tighten each screw half way, rotating between the two before tightening them all
the way.