
Soak the manifold tubes in cleaning fluid | 185
Soak the manifold tubes in cleaning fluid
Keeping the dispense and aspirate tubes in a wetted condition is required to keep
them clog free. Soak the tubing in deionized or distilled water or a cleaning fluid
whenever the device is not being used to ensure trouble-free performance.
Prime Step
1. Fill a supply bottle with a cleaning agent.
2. Click
W-Prime
or
S-Prime
and define parameters to fully prime the tubing, for
example:
Washer
Syringe Dispenser
Volume = 5000 µL
Volume = 8000 µL
Flow Rate = 5
Flow Rate = 5
Submerge tips
in fluid after prime: enable option and set
Duration
, up to 24
hours.
n
For the manifolds: Use the priming trough inserts to submerge the tips.
n
Note
: the aspirate manifold does not completely evacuate the priming trough
insert.
n
Overnight Maintenance
: You can turn off the instrument after the tips are
submerged to leave them soaking in the priming trough for an extended
duration.
3. Click
Run
. Save the program with a memorable name for future use.
AutoPrime
Enable
AutoPrime
to ensure the tubing is soaked during downtime intervals.
It is especially important when using the 32-tube SB dispense manifolds to keep
the tubes wet to prevent clogs. One option is to modify the AutoPrime parameters
used for idle periods during a regular work day to soak the tubing for longer
periods overnight and on the weekends.
n
Be sure to use the priming trough inserts when the submerge feature is
enabled. And, remove either the Peri-pump cassette or its priming trough
BioTek Instruments, Inc.