Parts Name
Replacement Cycle (approx.)
Part No.
How to Replace the Inner Lid Gasket
The Inner Lid Gasket replacement part is available from an authorized dealer (Part No: 0401026)
The Inner Lid Gasket may need to be replaced due to wear and tear. Check it once a year.
When the Inner Lid Gasket turns hard and white (the original
color is a semitransparent), cracks or tears, loses its pliancy,
or for any reasons which allow water (or steam) to escape
from the gap between the Upper Lid and the Body
Purchase and replace with a new one
(must be a manufacturer-approved replacement part)
Pry out and pull off the old Inner Lid Gasket from the outer
circumference. Then, fit the new Inner Lid Gasket securely to the
same position as shown in the figure.
Inner Lid Gasket
1
How to Clean with Citric Acid
The citric acid is available from an authorized dealer (Part No: C71502017-013)
Depends on the quality of water that is
added to the Pure Water Chamber.
Plug the product, ensure the product is ready and the Filter Chamber is filled with RO
water (see pg.09).
1
Put 20g of the Citric Acid Cleaner (P/N: C71502017-013) in a cup and add warm water to
dissolve first. Pour it into the Inner Container and fill with clean water to the MAX line.
2
Close the Upper Lid, then press “START” button for 10 seconds to activate the Cleaning
function.
3
Dispense the water to clean the water dispenser. Refill clean water in the Inner Container
to the MAX line and repeat the dispense operation as many times as needed until the smell
of the citric acid is removed throughly.
4
* Do not fill over the MAX line.
• Perform every 1-3 month(s) of usage.
• Do not drink the water used for citric acid cleaning.
• Only used for the Inner Container (NOT FOR THE PURE WATER CHAMBER).
* Repeated citric acid cleaning may be necessary for hard-to-remove residues or if water does not dispense
smoothly.
* Take precaution when disposing remaining water from the Inner Container. Be sure to unplug the unit, remove
the Drip Grid and Drip Tray, and drain all the water out of the the Drain Tube beforehand.
• Consumable parts are available from your authorized distributor/dealer.
• When replacing parts, please check the model number and the parts name beforehand.
Inner Lid Gasket
Filter Cartridge
Hydrogen Cell
0401026
C71502017-011
C71502017-012
12 months
6 - 12 months
24 - 36 months
* The cleaning time is 30 min.
Cleaning
30:00
Cleaning Done
Press any key
DISPLAY
DISPLAY
Replacement Parts
25
#1
#2
#1
#2
Depends on the frequency of usage and
water quality.