74
| Chapter 3: Operation
Note:
The reader stores measurement values for the last 8 plates in
memory.
Result
R E P O R T : H B S - A G
I
D
:
0
0
1
0 7 /
1 7
/
0
7
•
Use the
OPTION
key to select the appropriate
Plate ID
and
Report.
Note
that the
Assay ID
will change if the selected
Plate ID
was read with a
different assay. Once you have found the correct
Plate ID
, press
Enter
.
Editing Standard Outliers
If a standard curve was generated and if
EDIT STANDARD OUTLIERS
was set to
MANUAL
in the assay definition, the option to edit outliers is presented.
E D I T
S T D
O U T L I E R S :
Y E S N O
•
Select
NO
to include all standards in the curve-fit calculations.
•
Select
YES
to indicate that one or more standard replicates or groups
should be temporarily excluded from curve fit-calculations.
¾
If
AVERAGE STANDARDS
was set to
NO
in the assay definition, one or
more standard
replicates
can be chosen for exclusion.
E D I T S T D 1 R E P 1 ?
Y E S
Y E S N O
•
Select
YES
to exclude the replicate from curve-fit calculations.
•
Select
NO
to retain the replicate.
•
Press
ENTER
to advance to the next replicate.
¾
If
AVERAGE STANDARDS
was set to
YES
in the assay definition, one
or more standard
groups
can be chosen for exclusion.
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...