10: Install Software/Connect to Computer (Optional)
|
21
5.
PROMPT SAMPLE COUNT allows the user to enter the total number of samples
on the plate at runtime; only those sample results will be printed. Select
YES
to
present this prompt at run-time.
Read Speed
R E A D I N
R A P I D
M O D E ?
N O
Y E S N O
RAPID MODE
reads a 96-well plate at 6-second intervals with wavelengths higher
than 400 nm. Selecting NO means a plate with single-wavelength kinetic
measurements will be read in intervals less than 12 seconds (“Regular” or “Normal”
Mode). Specifications for these modes are outlined in
Chapter 1
.
10: Install Software/Connect to Computer
(Optional)
The ELx808 has serial (RS232) port located on back of the reader (see
This port allows the reader to communicate with a computer using the BioTek Gen5
software. It also allows for upgrades to the ELx808 basecode (on-board) software.
Attach the Cable
•
Turn off the computer and the reader.
•
Connect the supplied serial cable to both machines.
Install Gen5 Software on the Host Computer
If applicable, install Gen5 on the host computer. There is a
certain sequence of events that
must
be followed to ensure
that the software is properly installed and configured. Please
follow the instructions provided in
Gen5 Getting Started
Guide
to install the software.
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...