| Chapter 2: Installation
9: Configure Reader Settings
The ELx808 may be configured a number of ways, depending on user preference.
Configuration options are accessed via the Select Utility Option menu, and include:
•
SETUP
: Set the date and time.
•
OUTPUT
: Select whether reports will be output to a Printer, the Computer, or
both, choose the Column or Matrix report format, and determine if Curves will
be printed.
•
READ
: Indicate whether prompts shall be displayed at run-time for Plate IDs,
Sample IDs, and Sample Counts.
•
TESTS
: Access functions to test reader optics and instrument calibration.
Check the version of basecode software installed. See
Chapter 4,
Instrument Qualification
for more information.
To select the SETUP, OUTPUT, and READ user-configurable options:
•
At the Main Menu screen, select
UTIL
to access the Select Utility Option menu.
S E L E C T U T I L I T Y
O P T I O N :
T E S T S S E T U P
O U T P U T
R E A D
SETUP Options
1.
At the Select Utility Option screen, select
SETUP
.
E D I T S E T U P
I N F O R M A T I O N :
D A T E T I M E
F I L T E R
* M O R E
2.
At the Edit Setup Information screen, select
DATE
.
D A T E : 1 0 / 0 9 / 0 7
M D Y
M M D D Y Y D D M M Y Y
3.
Enter the new date, using the numeric keys. The cursor is positioned under the
first editable field, and advances automatically. To change the date format, select
MMDDYY
or
DDMMYY
. The display updates to reflect the new format.
4.
Press
Enter
to return to the Edit Setup Information screen.
T I M E : 0 3 : 1 1 P M
1 2 H R
1 2 H O U R 2 4 H O U R
A M / P M
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...