8: Check/Adjust the Reader’s Filter Table
|
17
8: Check/Adjust the Reader’s Filter Table
After installing the filter wheel (or new filters), ensure that the ELx808’s filter table
accurately maps the physical location of the filters in the filter wheel.
1.
Turn on the reader if it is not already on.
2.
At the Main Menu screen, select
UTIL
.
R E A D Y 0 1 : 3 0 P M 1 0 / 0 9 / 1 0
R E A D D E F I N E
R E P O R T
U T I L
The SELECT UTILITY OPTION menu appears:
S E L E C T U T I L I T Y
O P T I O N :
T E S T S S E T U P
O U T P U T
R E A D
3.
From the Select Utility Option menu, select
SETUP
. The EDIT SETUP
INFORMATION screen appears:
E D I T S E T U P
I N F O R M A T I O N :
D A T E T I M E
F I L T E R
* M O R E
4.
Select
FILTER
. The wavelength currently defined for Filter #1 appears:
E N T E R
F I L T E R # 1
W A V E L E N G T H : 4 0 5
If you need to change the filter wavelength number, use the numeric keypad to
enter a number at the cursor location. To save the entry and move to the next
filter on the filter table, press the
Enter
key. When the last filter has been
entered, the software exits the filter routine, and displays the following screen:
E D I T S E T U P
I N F O R M A T I O N :
D A T E T I M E
F I L T E R
* M O R E
5.
Press the
Main Menu
key to return to the Main Menu.
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...