16
| Chapter 2: Installation
Printers
The ELx808’s parallel port (LPT1) allows connection to compatible printers. The reader
supports printers using either HP's PCL3 language, such as the HP DeskJet series, or
Epson's LQ language. For a list of compatible printers call BioTek's Technical
Assistance Center or visit our website (refer to
Chapter 1
for contact information).
7: Turn on Reader and Run System Test
If all of the required steps preceding this one have been performed successfully, locate the
reader’s power switch (on the right side) and turn on the reader. The ELx808 will
automatically perform a System Test. The System Test conducts a series of tests at each
wavelength defined in the filter table to confirm adequate light levels, low electronic noise,
adequate photodiode sensitivity, overall system cleanliness, and (if equipped) proper
function of the incubator. The testing verifies that the reader will give in-specification
performance for each set wavelength over the specified OD range.
If a printer is connected to the reader, the test results automatically print. This test can also
be run through Gen5 running on a host PC; consult the Gen5 Help system for instructions.
•
If the test passes, the reader will beep once and the display will show the
software’s main menu, which will resemble one of the following:
Standard ELx808:
R E A D Y 0 1 : 3 0 P M 1 0 / 0 9 / 1 0
R E A D D E F I N E R E P O R T
U T I L
ELx808IU:
R E A D Y 0 1 : 3 0 P M 3 7 . 0 º C
R E A D
D E F I N E R E P O R T
U T I L
•
If the test fails
, the reader will beep repeatedly and the display will show an
error code. If this happens, write down the error code and then press the
STOP
key on the keypad to stop the beeping. Look up the error code in
Chapter 6
,
Error Codes
to determine its cause.
If the problem is something you can fix (for example, if the error code is 0201
or 0301, indicating that the filter wheel is missing), turn off the reader, fix the
problem, and then turn the reader back on. If the cause is not something you
can fix, contact BioTek’s Technical Assistance Center. See
Chapter 1,
Introduction
for contact information.
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...