| Chapter 4: Instrument Qualification
Do not remove the filters from the test plate, and do not
use alcohol or other cleaning agents to clean them.
¾
Check the microplate carrier to ensure that it is clear of debris.
•
Channel-to-Channel Variation
:
This test ensures that selected channels
read the same value for a filter as their paired channel when the plate is
“turned around” 180° in the plate carrier. It is a way to verify that each
channel’s Delta values are adjusted relative to the signal strength of the
reference channel.
Empty Carrier Test
The Empty Carrier Test confirms the ELx808’s read capabilities at the 100% light level,
and can help to identify deteriorating interference wavelength filters and other optical
problems.
Liquid Tests
Conducting Liquid Tests confirms the ELx808’s ability to perform to specification with
liquid samples. Liquid testing differs from testing with the Absorbance Test Plate in
that liquid in the wells has a meniscus, while the Test Plate’s neutral density filters do
not. The optics characteristics may differ in these two cases, alerting you to different
types of problems. The Absorbance Test Plate will indicate the absolute amount of light
absorbed, which will effectively test the linearity of the electronics. The liquid tests will
help detect optical defects such as dirt on the lenses or other contamination that can
contribute to errant readings.
•
Liquid Test 1
tests the alignment, accuracy, repeatability, and channel-to-
channel variability of the reader, making evident any problems with the
optics of the system. If you have the Absorbance Test Plate, you will only
need to run Liquid Test 1 for routine testing.
•
If you do not have the Absorbance Test Plate, test the linearity,
repeatability, and alignment of the reader by preparing a series of solutions
of varying absorbances as described in
Liquid Test 2
.
•
Liquid Test 3
is an optional test at 340 nm, offered for sites that must have
proof of linearity at wavelengths lower than those attainable with the
Absorbance Test Plate. This test is considered optional because the reader
has good “front end” linearity throughout its wavelength range.
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...