page 56
Chapter 4:
Operating Sally Sue and Peggy Sue
Sally Sue and Peggy Sue User Guide
If Running Buffer is present in the separation tray, click
Cancel
and remove the buffer manually. When Run-
ning Buffer evaporates, it leaves a highly viscous residue that the automatic cleaning feature can’t remove.
To remove Running Buffer:
1.
Select
Instrument
and click
Open Trays
.
2.
Click
Separation
to open the separation tray.
3.
Add 800 μL of deionized water to the troughs in the separation tray and let soak for 20 minutes.
4.
Remove the water by either aspirating with a pipette or with the instrument’s vacuum wand.
5.
Repeat the steps above until the Running Buffer or residues are completely removed.
6.
To complete the cleaning process, select
Instrument
and click
Cleanup
.
Self Test
Sally Sue and Peggy Sue can perform a set of self tests to make sure they’re operating properly. To start the
test, select
Instrument
and click
Self Test
. The test takes about two minutes.