Starting a Run: Size-based Assays (Sally Sue and Peggy Sue)
page 35
Sally Sue and Peggy Sue User Guide
5.
Place the assay plate into the sample tray.
6.
Click
Start
.
Step 2 - Start the Run
You can start a run two ways depending on whether you want to run an assay using existing parameters or
set up a new assay.
1.
To start a new run with an existing assay:
a. Select
File
in the main menu and click
Open Assay
.
b. A list of the last five assays opened will display. Select one of these assays or click
Browse
to open
the Assay folder and select a different one.
2.
To start a run with a new assay:
a. Select
File
in the main menu, click
Open Assay
and select
Sally Sue Size
or
Peggy Sue Size
. From
the pop up window, select the
Assay Type
: Size, Total Protein Size or Charge (Peggy Sue only) and
the
Size Range
(Size or Total Protein Size assay types only): 2-40 kDa, 12-230 kDa or 66-440 kDa.
b. The
Start
button will display which means an assay is already loaded.
Summary of Contents for ProteinSimple Peggy Sue