Auxiliary Module
page 29
Sally Sue and Peggy Sue User Guide
The
Start Run Wizard
in Compass for Simple Western will also check the fluid levels in each bottle. When a
full or empty status is detected, Compass will display an error in the
Start Run Wizard
:
If this happens, fill or empty each bottle using the procedures in the next section.
Removing and Installing Bottles
Bottles may be removed from the auxiliary module whenever Sally Sue or Peggy Sue are idle. To remove or
install the waste or water bottle, use the following procedure:
IMPORTANT
Don’t remove either bottle if Sally Sue or Peggy Sue are running, doing this may abort the run in
progress.
1.
Remove the bottle by unscrewing it clockwise until it disengages from the manifold.
2.
Pull the bottle down slowly until the sensor at the bottom of the bottle is clear.
3.
Fill the water bottle to the base of its neck and empty the waste bottle as needed.
Summary of Contents for ProteinSimple Peggy Sue
Page 12: ...page 8 Chapter 2 Getting Your Lab Ready Sally Sue and Peggy Sue User Guide ...
Page 36: ...page 32 Chapter 3 Sally Sue and Peggy Sue Sally Sue and Peggy Sue User Guide ...
Page 64: ...page 60 Chapter 4 Operating Sally Sue and Peggy Sue Sally Sue and Peggy Sue User Guide ...
Page 76: ...page 72 Chapter 5 Maintenance and Troubleshooting Sally Sue and Peggy Sue User Guide ...