page 28
Chapter 3:
Sally Sue and Peggy Sue
Sally Sue and Peggy Sue User Guide
Auxiliary Module
The auxiliary module holds the water and waste bottles used by Sally Sue and Peggy Sue. The waste bottle
has a yellow label, and the water bottle has a blue label. The volume in each reservoir is 1.0 L.
Deionized water is used by the instrument to clean the troughs in the separation tray and the fluidics in the
head assembly. The waste bottle collects waste generated during the run. For more information on waste
production, please refer to the “Waste Production and Disposal” on page 78.
There are two liquid level sensors in each bottle. The sensor at the bottom of the bottle detects when the
bottle is empty, and the sensor at the top of the bottle determines if sufficient volume is available to com-
plete an eight-cycle assay with the instrument.
NOTE: We recommend you check fluid levels in the bottles before starting a run. Empty the waste bottle if
needed, and fill the water bottle up to the base of its neck. If the waste bottle fills up or the water bottle
empties during a run, the run will be aborted.
Drip Tray
Manifold
Water Reservoir
Alignment Labels
Waste Reservoir
Alignment Labels
Waste Reservoir
Water Reservoir