FCS2 (Focht Live-Cell Chamber System) Instructions

A Message from the Developer:

The FCS2 utilizes the most advanced technology for optimum functionality and versatility for the observation of living cells on a 
light microscope.  As part of our dedication to respond to our customers needs we would be very appreciative of any comments or 
suggestions you may have relative to our products continued improvement.

Before You Begin:

• 

Check to make sure there is no visible damage in shipment and save all packing materials.

• 

The Controller box should contain; controller, power cord, 2.5mm screwdriver, 1/8” jack, and 3 pin mini DIN 
connector.

• 

Check the packing list and call your supplier if there are any discrepancies.

Front Panel of FCS2 & FCS3 Controller

Figure 1

Control Interface Descriptions:

Numeric Display 

should be illuminated when the controller is turned on.  If it does not light, check the main power source then the 

fuse block located in the power input module where the AC power cord plugs in.

Display Selector

 is the black button adjacent to the display.  It selects the source of the information displayed.  When the controller 

is turned on, the default condition is to display the setpoint value indicated by the red light adjacent to the word 

Setpoint

.  The button 

selects one of a sequence of display options.  This button can be pressed at any time and has no effect on the temperature regulation 
functions of the controller.

Display Lights

 indicate the source of the value displayed on the numeric display.

Setpoint

 is adjusted by a potentiometer accessible with the 2.5mm screwdriver provided.  To adjust the setpoint set the numeric 

display to 

Setpoint

, then adjust the potentiometer to the desired value.  If you need to change the setpoint value during your 

experiment on a routine basis there is a remote setpoint jack on the back of the controller.  See Remote Setpoint.

Slide 

is the location where the calibration of the circuit reading the Microaqueduct Slide can be adjusted.

Chamber 

is the location where the calibration of the circuit reading the Chamber can be adjusted.

Reset

 (Red Button) resets the controller if the alarm sounds.  Note:  If the alarm went off there must have been a reason.  Check the 

chamber before resetting the controller.

Calibrate

 substitutes precision resistors into the circuit that reads the thermistors so that calibration adjustments can be made to a 

known value.

Chamber 

 is a six pin mini DIN connector where the FCS2 or FCS3 is plugged in.

Summary of Contents for FCS2

Page 1: ... of the information displayed When the controller is turned on the default condition is to display the setpoint value indicated by the red light adjacent to the word Setpoint The button selects one of a sequence of display options This button can be pressed at any time and has no effect on the temperature regulation functions of the controller Display Lights indicate the source of the value displa...

Page 2: ...ogramming or cycling operating temperatures Controller Operation 1 Turn the unit on using the main on off switch located on the rear panel The Chamber need not be plugged in at this time All further adjustments are made on the front panel 2 Set the display switch to the SET position Insert the enclosed screwdriver into the hole marked Setpoint and adjust the controller to the desired setpoint 3 Se...

Page 3: ...e and press around the perimeter creating a seal Perfuse some media through the perfusion port producing a bead on the surface of the slide to displace air trapped in the line Lower the coverslip with cells onto the bead until it is resting on the gasket Place the closure assembly on top of this stack while aligning the black electrical connector with the oval slot in the chamber top Maintain a ge...

Page 4: ...ll This is due to the fact that the optical coupling medium oil glycerin or water also has a thermal coupling effect In this case a Bioptechs Inc Objective Heater and an Objective Heater Controller will be necessary for uniform temperature across the field Objective Heater Description The Bioptechs Objective Heater is an adjustable thin film size and shape compliant heating band and temperature se...