background image

© Biotage 2020

Quick Start Guide to TurboVap

®

 LV | Page 1

For detailed instructions, see the TurboVap

®

 User Manual.

Mode (1)

Display shows the current operation mode. Mode is controlled 
by pressing the blue button. There are three evaporation modes 
available on TurboVap LV:

 

»

Manual:

 Evaporate until the user presses the 

Stop

 button.

 

»

Time: 

Evaporate for a set period of time.

 

»

Method: 

There are several options. For more information, 

see “Define Methods” in the TurboVap

®

 User Manual.

Evaporate

 Settings

System Ready

12:13 Wed 22 Jan 2020

 Start  

 Stop 

Edit

Time

 Mode

 Nozzle Setup

1.2 

L/min

L/min

0.0

54 min

 Time

 Status

 Gas Flow

75 

°

C

°

C

75

 Bath Temp

54:00

Remaining (min:sec)

0:00

Elapsed (min:sec)

1

2

6

9

3

4

5

8

7

Time (2)

Display shows the set evaporation time in minutes for the 
current evaporation method. Time is controlled by pressing the 
blue button. (Only shown in Time mode.)

Gas Flow (3)

Display shows the current and set gas flow in L/min. Gas flow is 
controlled by pressing the blue button.

Bath Temperature (4)

Display shows the current and set water bath temperature in °C 
or °F. Bath temperature is controlled by pressing the blue button.

Nozzle Setup (5)

Display shows which nozzle rows are currently enabled. 
Rows are selected ON or OFF by pressing the blue button. 

Status (6)

Display shows the remaining and/or elapsed time of the 
current evaporation.

Quick Start Guide to TurboVap

®

 LV 

Start/Pause (7)

Press to start or pause the evaporation in progress.

Stop (8)

Press to stop the evaporation in progress.

Settings (9)

Press to define system settings. Displays an alternative screen.

Evaporate

 Settings

System Ready

16:48 Wed 22 Jan 2020

 Manifold Setup

48 Positions

 Sound Level

Medium

    Sleep Mode

Edit

 Temp Unit

°

C

 Brightness

-

+

Change

 Time & Date

16

48

:

2020

Jan

22

Bath Light

OFF

ON

More »

Lid Sleep

OFF

ON

Cont End-Point

OFF

ON

10

11

14

15

18

12

13

19

17

16

Sound Level (10)

Display shows the current sound level. Sound level is controlled 
by pressing the blue button. The sound level can be set to Low, 
Medium, High, or OFF.

Brightness (11)

Display shows the brightness of the touch screen. Brightness 
is controlled by pressing either 

+

 or 

.  

Manifold Setup (12)

Display shows the selected/installed manifold. For more 
information, see “Manifolds” in the TurboVap

®

 User Manual.

Sleep Mode (13)

Display shows the sleep mode settings. For more information, 
see “Sleep Mode and Automatic Wake-Up” in the TurboVap

®

 

User Manual.

Temp Unit (14)

Display shows the temperature unit used by the system.  
Press 

Change

 to switch between °C or °F.