background image

1. Turn on the InBody230 and keep the power on until the home screen

     initializes. It will take 15 seconds. During the warming, do not touch

     the base.

2. Step on the foot plate and measure weight. 

3. Enter your personal information and press “ENTER” button. 

Installation

Measurement

1. Raise the stand and lift the head and lower the arms.

2. Push the knob shown above and turn it toward right to tighten the joint screw. 

3. Plug the power adapter cable into the InBody230. 

4. If the air bubble is not in the center of the level indicator, tighten or loosen

   the leveling screws(4EA) on the foot plate to adjust the level of InBody230.

   InBody230 must be leveled properly for accurate weight measurement. 

BODY COMPOSITION ANALYZER

Install Manual 

1. If you use the InBody230 on a carpet, static electricity may cause serious damage to the
   electronic components. If the InBody230 is to be installed on a carpet, use antistatic mats
   below the InBody230. Use InBody tissue to wipe feet and hands before the measurement
   starts. 
2. Do not wipe the base and hand electrode with cleaning liquids. If cleaning liquids enter 
   into InBody230, it may cause corrosion or electric shock by short circuit. For cleaning of
   the InBody230, please use InBody tissue, which also has antibacterial functionality.
3.Do not dismantle InBody230. Internal parts are not for customer use and may cause
   electric shock. If InBody230 is dismantled, the warranty is void, and service costs will be 
   charged. 

DANGER

NOTE

CAUTION

Please keep the packaging for future events of repacking InBody230 and sending 

it for servicing or otherwise transporting InBody230 elsewhere.

Head

Stand  

Base

Hand Electrode

Input Buttons

Foot Electrode

Contents of the Box

Basic

InBody230

Install Manual 1EA

User’s Manual CD 1EA

Warranty

Adapter 1EA (12V, 3.5A)

Power Cable 1EA

Optional

Thermal printer

SD400 Set

Result Sheet

Carrying Bag

InBody Tissue

<Side view of InBody230>

< Leveled state >

Level Indicator

Leveling Screws

* Make sure the power cord is firmly plugged in.

<Rear view of InBody230>

<Rear view of InBody230>

Power switch

<Printer Cable / Option>

A

B

C

A

B

C

DEL

ETE

ABC

QZ

DEF

GHI

JK

PRS

TUV