Page 1: ......

Page 2: ...ducts or services we would like to hear from you Please send all feedback to Manufacturer SIA Biosan Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 674 261 37 Fax 371 674 281 01 https biosan lv Market...

Page 3: ...rs 10 7 Specifications 10 8 Ordering information 11 9 Care and maintenance 12 10 Troubleshooting errors 12 11 Warranty 14 12 EU Declaration of conformity 15 1 About this edition of the instructions 1...

Page 4: ...rer Save the unit from shocks and falling Store and transport the unit as described in section 10 Storage and transportation Use only original parts and accessories provided by manufacturer for this p...

Page 5: ...of the unit in specific at mospheres Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms Do not lean on the unit during operation Do not opera...

Page 6: ...reparation for subsequent academic pharmaceutical biotechnological or biomedical studies Homogenizer facilitates the formation of a supernatant containing nucleic acids and proteins suitable for subse...

Page 7: ...ling must be carried out by two persons 4 2 Complete set Package contents RCP 24 Reciprocal Laboratory Homogenizer 1 pce Power cable 1 pce Rotor stand 1 pce Emergency lid opening tool located at the r...

Page 8: ...ressing the open fig 2 8 Display shows the OPEN indication 5 2 3 Unscrew the rotor lid fig 1 3 remove it then load the samples into the rotor fig 1 2 Caution Insert the first 2 tubes opposite to each...

Page 9: ...nt by pressing the RUN STOP key Timer resets to the starting values 5 5 Power interruption In case of mains power failure the unit is not equipped with automatic restart function which requires the us...

Page 10: ...Glass 0 1 mm 2 5 g cm3 2000 Microbes yeast Glass 0 5 mm 2 5 g cm3 Animal soft tissues muscle kidney heart spleen etc Ceramic 2 8 mm 6 g cm3 2000 Leaves Stems Roots Ceramic 2 8 mm 6 g cm3 2000 Hard an...

Page 11: ...programme of improvement and reserves the right to alter design and specifications of the equipment without additional notice Test tube rotor capacity up to 24 Test tubes 2 ml Speed control range 500...

Page 12: ...it shakes when in operation There is no rotor inside the unit or the first two tubes are not loaded symmetrically Stop the centrifuge and load the first two tubes symmetrically re start the unit Displ...

Page 13: ...A type M time lag Medium rating 250 V Figure 4 Fuse holder 9 4 Disposal Disposal of the appliance requires special precautions and must be carried out at an appropriate disposal site separate from nor...

Page 14: ...certified and sent to the local distributor address To obtain the claim form visit Technical support page on our website at link below 11 5 Extended warranty For RCP 24 the Basic Plus class model ext...

Page 15: ...ular requirements for laboratory equip ment for mixing and stirring EMC 2014 30 EU LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General requirem...

Page 16: ...Biosan SIA Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 https biosan lv Edition 1 06 July of 2022...