Summary of Contents for Inteliwasher 3D-IW8

Page 1: ......

Page 2: ...wn 14 7 Program error messages 15 8 Specifications 16 9 Ordering information 17 10 Care and maintenance 18 11 Warranty 21 12 Glossary 22 13 EU declaration of conformity 23 1 About this edition of user...

Page 3: ...sults depends on the quality of the microplate rins ing procedure To assure correctness of the rinsing procedure regularly exercise vis ual control over reproducibility of liquid volume dosing Before...

Page 4: ...aste bottle to overflow the maximum level during operation there is a mark on the side surface of the bottle Take necessary precautions utilizing waste liquid in accordance with general laboratory sta...

Page 5: ...plate with rectangular wells Note Optional 4 channel washing solution weight logger 4CHW Logger al lows automatic buffer bottles and waste bottle volume control The unit is supplied with 8 channel was...

Page 6: ...600 mm 5 pieces 7 Tube for manifold outside inside diam length 3 2 1 6 400 mm 1 piece 8 Tube for manifold outside inside diam length 5 3 440 mm 1 piece 9 Tube for hydrophobic filter outside inside dia...

Page 7: ...the unit To provide optimal ventilation ensure 100 mm clearance on each side Opera tion conditions are described in the Specification section 4 4 2 Remove the protecting cover 4 4 3 Unpack the plate p...

Page 8: ...be fig 1 13 Check that the IN marking on the filter is facing the bottle 5 fig 11 4 4 11 Connect the Waste bottle 4 to the half litre bottle 5 fig 3 10 4 4 12 On the rear side of the unit 4 4 13 Conne...

Page 9: ...fig 3 8 on the rear panel of the unit 4 5 2 Connect the power cord to the external power supply fig 1 16 4 5 3 Connect the external power supply to the grounded power socket 4 5 4 Switch on the unit...

Page 10: ...s Categories FastFRAME 2 4 and FastFRAME C 5 are intended for Watman Schleicher Shuell FastFRAME plates with rectangular wells Program 01 02 03 04 05 06 07 08 09 10 Dispense NO YES YES YES YES YES YES...

Page 11: ...different type of plate is installed perform plate setup to adjust the manifold lowering depth see paragraph 4 20 Note When working with the FastFRAME plates remove the plate holder fig 2 2 Install i...

Page 12: ...g or shaking then the elapsed time is displayed 5 17 If 4 channel washing solution weight logger is connected to the unit then during op eration display shows program number currently used bottle numb...

Page 13: ...of different type is used setup the manifold needles lowering depth Press the Enter key in the start mode the following message appears on the display fig 10 To calibrate press Run Stop key The manif...

Page 14: ...sing and fig 11 2 Press Run Stop fig 11 3 key and the system will perform rinsing cycle Repeat the procedure twice if necessary 6 3 Disconnect the tube from selected bottle and press Prime rinse key a...

Page 15: ...n CANCELED BY OPERATOR Program execution is stopped by operator Press Run Stop key BOTTLE 4 OVERFLOWED Overflow of waste collection bottle Empty the bottle BOTTLE EMPTY Buffer bottle needs to be fille...

Page 16: ...e increment 25 l Irregular liquid dosage at 300 l max 2 5 or 7 5 l Allowed residual liquid volume in plate well max 2 l Number of wells washed simultaneously 8 Number of washing cycles 1 15 Aspiration...

Page 17: ...er max loading per scale cup 2 kg dimensions 267x252x97 mm weight1 3 kg BS 060102 AAI 9 3 Replacement parts Replacement Description Catalogue number Bottle 1 Assembled with weight tube filter BS 06010...

Page 18: ...the rinsing cycle twice using distilled water after finishing the operation see section 5 10 4 2 Monthly With 75 ethanol solution clean mounting surface of the plate platform and plate holder spring 1...

Page 19: ...caused by needles they may contain dangerous infections Take off the manifold from the device put it down with needles upward Clean the aspiration channel needles with the wiping rod 0 80 mm and clean...

Page 20: ...t the service engineer 10 8 Storage and transportation 10 8 1 Remove the manifold tubes and platform for plates 10 8 2 Put on packaging blocks onto sides of the unit and place it into the original car...

Page 21: ...or address To obtain the claim form visit Technical support page on our website at link below 11 5 Extended warranty For 3D IW8 the Premium class model one year of extended warranty is available free...

Page 22: ...which provides solution aspiration conse quently from two opposite points of a well Strip washing mode plate washing by rows Dispensing filling wells with a set solution volume Waste bottle the bottl...

Page 23: ...ical equipment for measurement control and laboratory use General requirements EMC 2014 30 EU LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Gener...

Page 24: ...Biosan SIA Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 3 02 December of 2020...