Edition 1.-4.03
17
9.5
Troubleshooting
.
Problem
Possible reason
Troubleshooting
Unit not working
No power
Check external power supply and con-
nector
(
FTA-2i
) Trapping flask full
Sensor shut down the motor. Empty the
trapping flask according to
5.6
.
(
FTA-2i
) Sensor not connected
properly
Check sensor cable and attachments
Aspiration slow or
absent
(
FTA-2i
) Aspiration intensity set to
small
Turn the regulator knob clockwise to in-
crease aspiration.
Breach in the system
Check all the tubing and fittings. Tightly
close the lid of the unit
Dirt in the tubing
Clean the tubing, see
8.3
.
Liquid in the filter
Replace the filter
(
FTA-2i
) Sensor
activated, but the
flask is not full
Foam in the trapping flask activated
the sensor
Do not use the unit with a foaming liquid
or use antifoaming agent.
9.6
Disposal
. Disposal of the appliance requires special precautions and must be carried
out at an appropriate disposal site, separate from normal household waste. To prevent
pollution of the environment, all waste resulting from the disposal of the product must
be collected and disposed of in the country of use, in accordance with the applicable
requirements for the handling of electronic waste.
10. Storage and transportation
10.1 Store and transport the unit in a horizontal position (see package label) at ambient
temperatures between -
20°C and +60°C and maximum relative humidity of 80%.
10.2 After transportation or storage and before connecting it to the electric circuit, keep the
unit under room temperature for 2-3 hrs.
10.3 For extended storage, disassemble and clean according to
9.3
. Empty the trap flask.
The unit does not require special procedures.