Biorep Technologies PER-5.2 User Manual Download Page 5

 

 

Page 5 of 21 

1.5 

Site Preparation 

This machine is intended to be used  in a  controlled  laboratory  environment.  Please  refer to  the 
figure  below  for proper installation of the machine. The machine has  one  ventilation opening  in 
the rear. A 15 inches clearance is recommended between the back of the machine and the wall. 

 

Figure 1: Dimensions of the Perfusion Apparatus 5 

 

1.6 

Unpacking Instructions 

 

Remove the top 2 pieces  of blue foam and the items that are located on top of the 
machine. Then, carefully remove the foam that surrounds the machine.  
It is suggested that the machine be removed from the box by two people, one on 
each side. To lift the Perfusion Apparatus, place your hands underneath the 

          machine and remember to bend at the knee.  Please, keep the box and all the foam 
          pieces in storage, in case the machine needs to be shipped in the future. 

1.7 

Packing List 

The following items are included with the purchase of a Perfusion Apparatus: 
 

Qty 

Reference 

Description 

PER-05-115 

Perfusion Apparatus 

PER05-TUBHC-01 

Disposable Perfusion Tubing Set with Coil (Sterile) 

PERTRAY-MABS-01 

Disposable Perfusion Tray (Sterile) 

TC-02 

Biorep Thermoprobe 

86610610 

Medical-Rated Power Cord 

PER05-UM 

User Manual 

PER05-DAQ 

CD with Software & User Manuals 

 

 
 
 
 

 

15 Inches

Summary of Contents for PER-5.2

Page 1: ...Page 1 of 21 PERFUSION APPARATUS MODEL No PER 5 2 ______________________________________________________ USER MANUAL ______________________________________________________...

Page 2: ...PERFUSION TRAY 9 4 2 DISPOSABLE TUBING SET WITH COIL PER05 TUBHC 01 10 4 3 BIOREP THERMOPROBE TC 01 10 5 PERFUSION MACHINE SET UP 11 5 1 TUBING SET UP 11 5 2 HEATER SET UP 13 6 PERFUSION APPARATUS OPE...

Page 3: ...the literature that accompanies the device This symbol warns of moving parts may cause personal injury This symbol is used when two people are needed to do the task 1 2 Machine Information In the spa...

Page 4: ...accordance with the manufacturer s instructions See Site Preparation in section 1 5 for details 3 Do not defeat the safety purpose of the polarized or grounding type plug A polarized plug has two blad...

Page 5: ...s the machine It is suggested that the machine be removed from the box by two people one on each side To lift the Perfusion Apparatus place your hands underneath the machine and remember to bend at th...

Page 6: ...enables the researcher not only to analyze each perfusion but to refine the procedure into a repeatable and efficient protocol by creating custom recipes for the machine to execute automatically The a...

Page 7: ...tification 3 1 Front View Figure 3 Front View Touch Screen Automatic Heater Infrared Thermocouple Pressure sensor B Pump A Pressure sensor A Tray t c Input Pump B Transducer Protector Holder Transduce...

Page 8: ...Page 8 of 21 3 2 Rear View Figure 4 Rear View 3 3 Right View Figure 5 Right Side View ID Plate Vent USB Port 1 USB Port 2 LAN Port 1 Power Entry Module LAN Port 2...

Page 9: ...he opposite side are available to re circulate cold water or a suitable coolant to cool the enzyme if desired The Perfusion Tray is available in disposable and autoclavable versions The included dispo...

Page 10: ...g remains inside a closed peristaltic pump head for extended periods there may be permanent tubing profile deformation 4 3 Biorep Thermoprobe TC 02 The Biorep Thermoprobe is a sterile disposable type...

Page 11: ...he heating coil inside the automatic heater Make sure there is good contact between the stainless steel coil and the aluminum arms Also make sure there is good contact between the coil and the spring...

Page 12: ...the inlet or outlet 4 Connect Tubes A and B to the appropriate pressure sensor fittings 5 Secure both Transducer Protector A and Transducer Protector B to their respective holders 6 Use both Luer conn...

Page 13: ...ement s ends into the contact grooves The heating element terminations A and B are different Identify each one by their shape Be sure that end B of the heating element is inserted into the contact gro...

Page 14: ...beneath the covers can cause injuries ICON BUTTON CONVENTIONS The touch screen interface of this machine has virtual buttons represented by icons to activate and deactivate the different features of...

Page 15: ...some frequently used fields during the experiment process and during reporting The first step is creating a user account To create a new user account press the Add User Button 6 3 1 Add Users When th...

Page 16: ...Software Version installed in your perfusion machine and the machine Status Bar at the bottom The Status Bar will display any alarms or important messages related to the function of the machine You ca...

Page 17: ...that is editable by clicking These are usually variables set by the user to control the machine 6 5 2 1 Temperature Control The temperature control section of the control panel has the following func...

Page 18: ...ecified pressure in the loop The loop pressure is constantly measured and displayed by the Pressure PV indicator In Auto mode the Flowrate field on the bottom becomes an indicator of the current flow...

Page 19: ...Click on an experiment to select it All data displayed on the following tabs will relate to the selected experiment You can export all experiments of a particular user to an external device by selecti...

Page 20: ...he beginning of the next step and so on The time completion bar at the bottom shows you the experiment elapsed time in relation to the total time of the experiment To begin replaying the selected expe...

Page 21: ...it doesn t detect a closed cover To prevent exposure of hot surfaces make sure to close the cover properly before activating the heater 2 Heater enabled Pump off no flow This alarm is triggered when...

Reviews: