USER MANUAL
TD064-O&M-001
REVISION 4
Page 33 of 38
9
SPECIFICATION
Unit Description
Bioquell “Walk-in Chamber”
(WIC-1)
Unit Part Number
230V:
TD064-0000
Materials of Construction:
Stainless Steel, Phenolic Resin,
Glass
Temperature
Operating:
15
C - 30
C
(for successful bio
decontamination)
Relative Humidity
Operating:
5% - 80% RH
Power Requirements:
AC 230V, 50/60 Hz, 30A supply
Internal Dimensions
(WxHxD)
1680 (or 2180) x 2200 x 2320
(or 2720, 3120) mm
66.1 (or 85.8) x 86.6 x 91.3 (or
107, 123) inches
Doorway Opening Size
(WxH)
1100 x 2200 mm (43.3 x 86.6
inches)
Control System
Siemens S7 200 PLC
Ceiling Aeration Unit:
Airflow Rates
Out of cycle:
During Aeration:
140 m
3
/hr (82 cfm)
2200 m
3
/hr (1295 cfm)
Stirrer Fan Airflow
(per fan)
410 m
3
/hr (241 cfm)
Efficiency of HEPA filter FE1
(Aeration 2000 unit)
99.995% (H14)
Efficiency of HEPA filter FE4
(APB)
99.995% (H14)
Noise
50 dB(A)
*
(in aeration)
*
Measured outside a 1.7 x 2.6 x 2.5m stainless steel chamber.